Philips Medical Systems: Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)

Recall number
Z-0341-05
Recalling firm
Philips Medical Systems
Product
Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)
Status
Terminated
Initiated
2004-12-15
Posted
2004-12-21
Terminated
2016-02-02
Reason
In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy
Root cause
Other
Action
Philips notified customers on 12/13/04 by letter. Users are provided with action required to Mitigate Risk' until the correction is made by a Systems representative.
Quantity
2520 units
Distribution
Nationwide Foreign:Canada, Austria, Australlia, Belgium, China, Denmark, Finland, France, Germany, Israel, Italy,Kuwait, Latvia, Lebannon, Libya, Netherland,Norway, Portugal, Qatar, Russia, Singapore,South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand,Tunisia, UK, United Arab Emirates, Yemen
Lot, serial or model codes
Software versions A.02.00 and lower
510(k) numbers
["K031187"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register