Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter Lineo Opticap Disconnect Cap, product codes R5C4599Q (English) and N5C4599Q (dual English/French label); Baxter Healthcare Corporation, Deerfield, IL 60015 USA; 30 units per case

Recall number
Z-0341-04
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Lineo Opticap Disconnect Cap, product codes R5C4599Q (English) and N5C4599Q (dual English/French label); Baxter Healthcare Corporation, Deerfield, IL 60015 USA; 30 units per case
Status
Terminated
Initiated
2003-12-05
Posted
2004-07-20
Terminated
2004-07-30
Reason
Reports of loosening of the connections and disconnection of the device.
Root cause
Other
Action
Baxter sent recall letters dated 12/5/03 to the accounts in the United Kingdom and recall letters dated 12/8/03 to the accounts in Canada, informing them of the reports of loosening of the connections and disconnection of the device, and requesting that they discontinue use of product codes involved and revert the enrolled patients back to their traditional connection system.
Quantity
4,920 units
Distribution
Canada and the United Kingdom
Lot, serial or model codes
Product code R5C4599Q, lots H03J02029 and H03J27067 and product code N5C4599Q, lots H03I18083, H03I23018 and H03J29022
510(k) numbers
["K031676"]
PMA numbers
not on file
Product code
KGZ
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register