Skip to the content

FDA Medical Device Recalls (openFDA)

Aizu Olympus Co., Ltd.: Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.

Recall number
Z-0340-2024
Recalling firm
Aizu Olympus Co., Ltd.
Product
Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.
Status
Open, Classified
Initiated
2023-09-25
Posted
2023-11-17
Terminated
not on file
Reason
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Root cause
Under Investigation by firm
Action
Olympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Olympus is in the process of implementing a validated drying process at our repair centers. In the meantime, Olympus has established interim processes to perform an additional inspection of your scope for water prior to return. 1. Ensure your users are knowledgeable on the content of this letter. 2. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at olympus2806@sedgwick.com or by fax at 844-294-7184. For any questions about the acknowledgement form, please call the Sedgwick team at 888-667-1504. 3. Olympus requests that you report complaints, including infections, and endoscopes returned wet from our repair centers to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you require additional information, please do not hesitate to contact me at (647) 999-3203 or Cynthia.Ow@Olympus.com.
Quantity
33 units
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
Model Numbers: MAF-DM2 (04953170407260), MAF-GM (04953170340369), MAF-GM2 (04953170407208), MAF-TM (04953170288630), MAF-TM2 (04953170407147).
510(k) numbers
["K082720","K201300"]
PMA numbers
not on file
Product code
EOQ
Firm FEI
3002964398
Firm city
Aizuwakamatsu
Firm state
3 Chome
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register