FDA Medical Device Recalls (openFDA)

Caire, Inc.: CAIRE Liberator 45, MODEL 13262253, Liquid Oxygen System Unit

Recall number
Z-0337-2023
Recalling firm
Caire, Inc.
Product
CAIRE Liberator 45, MODEL 13262253, Liquid Oxygen System Unit
Status
Terminated
Initiated
2022-10-10
Posted
2022-12-02
Terminated
2024-08-19
Reason
An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
Root cause
Process control
Action
The firm issued an URGENT: MEDICAL DEVICE RECALL notice its consignees be email and telephone. The notice explained the issue and requested the removal of the affected products from the field and return to CAIRE. CAIRE will immediately replace the above serial numbers with new Liberator devices. Please contact CAIRE Customer Service via phone at 1-800-482-2473 or 770-721-7700 or via email at customerservice.usa@caireinc.com to arrange for the return and replacement of the above serial numbers.
Quantity
70 devices
Distribution
US Nationwide - Worldwide Distribution: CA, NC, OH, AZ, and Canada, Chile, Colombia, and Germany
Lot, serial or model codes
UDI/DI M766132622530, Serial Numbers: CBB3022311044, CBB3022311042, CBB3022311043, CBB3022311040, CBB3022311242, CBB3022311098, CBB3022311093, CBB3022311094, CBB3022311077, CBB3022311053, CBB3022311049, CBB3022311065, CBB3022311051, CBB3022311075, CBB3022311078, CBB3022311097, CBB3022311096, CBB3022311079, CBB3022311045, CBB3022311046, CBB3022311236, CBB3022311230, CBB3022311217, CBB3022311231, CBB3022311164, CBB3022311225, CBB3022311226, CBB3022311133, CBB3022311192, CBB3022311163, CBB3022311216, CBB3022311228, CBB3022311227, CBB3022311162, CBB3022311198, CBB3022311211, CBB3022311202, CBB3022311155, CBB3022311135, CBB3022311171, CBB3022311121, CBB3022311106, CBB3022311109, CBB3022311108, CBB3022311084, CBB3022311083, CBB3022311092, CBB3022311081, CBB3022311082, CBB3022311141, CBB3022311131, CBB3022311136, CBB3022311183, CBB3022311178, CBB3022311130, CBB3022311179, CBB3022311139, CBB3022311177, CBB3022311105, CBB3022311146, CBB3022311172, CBB3022311140, CBB3022311125, CBB3022311169, CBB3022311068, CBB3022311176, CBB3022311170, CBB3022311180, CBB3022311057, CBB3022311129
510(k) numbers
["K800742"]
PMA numbers
not on file
Product code
BYJ
Firm FEI
3004972304
Firm city
Ball Ground
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register