Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Vascular: RX ACCULINK Carotid Stent System 9 X 20mm, Part Number: 1011341-20, Lot Number: 0071361 The RX Acculink Carotid Stent System, used in conjunction with Abbott Vascular RX Accunet embolic protection system, is indicated for the treatment of patients at high risk for adverse events from carotid endarterectomy who require carotid revascularization and meet the criteria outlined below. (1) Patients with neurological symptoms and >50% stenosis of the common or internal carotid artery by ultrasound or angiogram OR patients without neurological symptoms and >80% stenosis of the common or internal carotid artery by ultrasound or angiogram, AND (2) Patients must have a reference vessel diameter within the range of 4.0 mm and 9.0 mm at the target lesion.

Recall number
Z-0337-2011
Recalling firm
Abbott Vascular
Product
RX ACCULINK Carotid Stent System 9 X 20mm, Part Number: 1011341-20, Lot Number: 0071361 The RX Acculink Carotid Stent System, used in conjunction with Abbott Vascular RX Accunet embolic protection system, is indicated for the treatment of patients at high risk for adverse events from carotid endarterectomy who require carotid revascularization and meet the criteria outlined below. (1) Patients with neurological symptoms and >50% stenosis of the common or internal carotid artery by ultrasound or angiogram OR patients without neurological symptoms and >80% stenosis of the common or internal carotid artery by ultrasound or angiogram, AND (2) Patients must have a reference vessel diameter within the range of 4.0 mm and 9.0 mm at the target lesion.
Status
Terminated
Initiated
2010-09-10
Posted
2010-11-10
Terminated
2010-11-16
Reason
The recall was initiated because Abbott vascular has discovered through internal testing that the affected lots may not meet the firm's quality specifications for catheter shaft tensile strength. If the affected device is used it may result in acute stent deployment difficulties and subsequent intervention.
Root cause
Pending
Action
Abbott Vascular sent an URGENT DEVICE RECALL letter dated September 10, 2010, to all consignees. The letter identified the product, the problem, and the action to be taken. The Sales Representatives were to personally visit each account to deliver the recall letter and assist in identifying and returning any unused devices to Abbott Vascular. Also, Abbott Vascular Regulatory Compliance initiated contact with the Risk Management department at each account by phone to verify their mailing address and provide a courtesy copy of the recall letter. The consignees were instructed to work with their local account representative to review their inventory, complete the Recall Effectiveness Check and return identified products to Abbott Vascular. For questions regarding this recall call (951) 914-3324.
Quantity
42 units total for all products
Distribution
Nationwide Distribution including AL, CA, CT, DC, FL, GA, IA, IL, IN, MA, ME, MI, MN, MO, OH, OK PA, TX, VA, and WI
Lot, serial or model codes
Lot Number: 0071361
510(k) numbers
not on file
PMA numbers
["P040012"]
Product code
NIM
Firm FEI
2024168
Firm city
Temecula
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register