Acute Innovations Llc: 10mm x 2.7mm RibLoc Screw, a component of the RibLoc Rib Fracture Plating System, Model RBP1202. The product is labeled in parts: "***Ribloc Rib Fracture Plating System***MATERIAL: Titanium***Rx ONLY***NON STERILE***SCREW***ACUTE Innovations LLC 21421 NW Jacobson Road, Suite 700 Hillsboro, OR 97124 USA www.acuteinnovations.com*** MediMark Europe Sarl*** 11 rue Emile ZOLA BP 2332****38003 GRENOBLE CEDEX 2 FRANCE" The Acumed Rib Congruent Bone Plate System is indicated for use in providing fixation during fractures, fusions, and osteotomies for the rib. The Acumed Rib Congruent Bone Plate System includes plates and screws designed specifically for the rib.
- Recall number
- Z-0336-2012
- Recalling firm
- Acute Innovations Llc
- Product
- 10mm x 2.7mm RibLoc Screw, a component of the RibLoc Rib Fracture Plating System, Model RBP1202. The product is labeled in parts: "***Ribloc Rib Fracture Plating System***MATERIAL: Titanium***Rx ONLY***NON STERILE***SCREW***ACUTE Innovations LLC 21421 NW Jacobson Road, Suite 700 Hillsboro, OR 97124 USA www.acuteinnovations.com*** MediMark Europe Sarl*** 11 rue Emile ZOLA BP 2332****38003 GRENOBLE CEDEX 2 FRANCE" The Acumed Rib Congruent Bone Plate System is indicated for use in providing fixation during fractures, fusions, and osteotomies for the rib. The Acumed Rib Congruent Bone Plate System includes plates and screws designed specifically for the rib.
- Status
- Terminated
- Initiated
- 2011-02-28
- Posted
- 2011-12-01
- Terminated
- 2011-12-06
- Reason
- RibLoc screw from lot L1008007 was out of specification. It would not engage with the hex tip properly.
- Root cause
- Process control
- Action
- The firm, ACUTE Innovation, sent an "URGENT NOTICE: DEVICE RECALL" letter dated February 21, 2011, to all consignees/customers via FED-EX on February 28, 2011. The letter describes the product, problem and actions to be taken. The customers were instructed to identify product having the lot code(s) and quarantine them; return all product from this lot to Acute Innovations immediately using shipping labels provided or return to: Acute Innovations, LLC, 21421 NW Jacobson Rd, Suite 700, Hillsboro, OR, 97124, and if they further distributed this product, notify those customers of this recall and request return of product(s). Customers can call ACUTE Innovation at 1-866-623-4137 for any questions about this recall.
- Quantity
- 299 units (267 in US and 32 to South Africa)
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, GA, IL, MN, NC, NY, UT, and WA; and countries of: South Africa.
- Lot, serial or model codes
- lot L1008007; Part # RBP1202
- 510(k) numbers
- ["K051410"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 3005670412
- Firm city
- Hillsboro
- Firm state
- OR
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28