Skip to the content

FDA Medical Device Recalls (openFDA)

Acute Innovations Llc: 10mm x 2.7mm RibLoc Screw, a component of the RibLoc Rib Fracture Plating System, Model RBP1202. The product is labeled in parts: "***Ribloc Rib Fracture Plating System***MATERIAL: Titanium***Rx ONLY***NON STERILE***SCREW***ACUTE Innovations LLC 21421 NW Jacobson Road, Suite 700 Hillsboro, OR 97124 USA www.acuteinnovations.com*** MediMark Europe Sarl*** 11 rue Emile ZOLA BP 2332****38003 GRENOBLE CEDEX 2 FRANCE" The Acumed Rib Congruent Bone Plate System is indicated for use in providing fixation during fractures, fusions, and osteotomies for the rib. The Acumed Rib Congruent Bone Plate System includes plates and screws designed specifically for the rib.

Recall number
Z-0336-2012
Recalling firm
Acute Innovations Llc
Product
10mm x 2.7mm RibLoc Screw, a component of the RibLoc Rib Fracture Plating System, Model RBP1202. The product is labeled in parts: "***Ribloc Rib Fracture Plating System***MATERIAL: Titanium***Rx ONLY***NON STERILE***SCREW***ACUTE Innovations LLC 21421 NW Jacobson Road, Suite 700 Hillsboro, OR 97124 USA www.acuteinnovations.com*** MediMark Europe Sarl*** 11 rue Emile ZOLA BP 2332****38003 GRENOBLE CEDEX 2 FRANCE" The Acumed Rib Congruent Bone Plate System is indicated for use in providing fixation during fractures, fusions, and osteotomies for the rib. The Acumed Rib Congruent Bone Plate System includes plates and screws designed specifically for the rib.
Status
Terminated
Initiated
2011-02-28
Posted
2011-12-01
Terminated
2011-12-06
Reason
RibLoc screw from lot L1008007 was out of specification. It would not engage with the hex tip properly.
Root cause
Process control
Action
The firm, ACUTE Innovation, sent an "URGENT NOTICE: DEVICE RECALL" letter dated February 21, 2011, to all consignees/customers via FED-EX on February 28, 2011. The letter describes the product, problem and actions to be taken. The customers were instructed to identify product having the lot code(s) and quarantine them; return all product from this lot to Acute Innovations immediately using shipping labels provided or return to: Acute Innovations, LLC, 21421 NW Jacobson Rd, Suite 700, Hillsboro, OR, 97124, and if they further distributed this product, notify those customers of this recall and request return of product(s). Customers can call ACUTE Innovation at 1-866-623-4137 for any questions about this recall.
Quantity
299 units (267 in US and 32 to South Africa)
Distribution
Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, GA, IL, MN, NC, NY, UT, and WA; and countries of: South Africa.
Lot, serial or model codes
lot L1008007; Part # RBP1202
510(k) numbers
["K051410"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3005670412
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register