Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1 Control Unit; Catalog number 801046.

Recall number
Z-0335-2007
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Advanced Perfusion System 1 Control Unit; Catalog number 801046.
Status
Terminated
Initiated
2004-10-12
Posted
2006-12-30
Terminated
2007-02-01
Reason
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
Root cause
Other
Action
U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Quantity
4,740 for all affected products
Distribution
Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Lot, serial or model codes
Serial numbers 0020 through 0025, 0030 through 0036, 0038 through 0040, 0042, 0044, 0045, 0047 through 0081, 0083, 0084, 0086 through 0102, 0106 through 0110, 0112 through 0130, 0132 through 0137, 0139 through 0147, 0149 through 0154, 0157 through 0166, 0169 through 0175 and 0177 through 0189.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register