Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: 3D TOP Ceiling Stand. X-Ray System model number 8773673

Recall number
Z-0335-06
Recalling firm
Siemens Medical Solutions USA, Inc
Product
3D TOP Ceiling Stand. X-Ray System model number 8773673
Status
Terminated
Initiated
2005-11-30
Posted
2005-12-31
Terminated
2006-07-19
Reason
screws on the rollers may become loose
Root cause
Other
Action
The recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.
Quantity
460 units
Distribution
The products were shipped to hospitals and medical facilities nationwide.
Lot, serial or model codes
Model number 8773673
510(k) numbers
["K802200"]
PMA numbers
not on file
Product code
IYB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register