FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: 3D TOP Ceiling Stand. X-Ray System model number 8773673
- Recall number
- Z-0335-06
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- 3D TOP Ceiling Stand. X-Ray System model number 8773673
- Status
- Terminated
- Initiated
- 2005-11-30
- Posted
- 2005-12-31
- Terminated
- 2006-07-19
- Reason
- screws on the rollers may become loose
- Root cause
- Other
- Action
- The recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.
- Quantity
- 460 units
- Distribution
- The products were shipped to hospitals and medical facilities nationwide.
- Lot, serial or model codes
- Model number 8773673
- 510(k) numbers
- ["K802200"]
- PMA numbers
- not on file
- Product code
- IYB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28