Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1 Temperature Monitoring System; Catalog number 802114.

Recall number
Z-0332-2007
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Advanced Perfusion System 1 Temperature Monitoring System; Catalog number 802114.
Status
Terminated
Initiated
2004-10-12
Posted
2006-12-30
Terminated
2007-02-01
Reason
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
Root cause
Other
Action
U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Quantity
4,740 for all affected products
Distribution
Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Lot, serial or model codes
Serial numbers 00019 through 00034, 00036, 00037, 00046 through 00053, 00055 through 00069, 00073 through 00114, 00116 through 00128, 00131 through 00193, 00195 through 00203, 00205 through 00334, 00339 through 00355, 00357 through 00414, 00416 through 00431. 00434, 00435, 00439 through 00447, 00448 through 00466, 00468 through 00473, 00475 through 00477, 00479 through 00582, 00584 through 00590, 00592 through 00594 and 00606.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register