Toshiba American Medical Systems Inc: Model CAT-850B Catheterization Table, a component of the Toshiba INFX-8000 series X-Ray Interventional System. These systems are single or dual plane systems that employ x-rays to obtain fluoroscopy or radiography images of the body. The system is comprised of a computer system, support arms (that contain the tube and solid state detector), x-ray generator and a patient table. This device is a digital x-ray system configured for use as a diagnostic x-ray angiography system. This x-ray angiography system is used for diagnostic and interventional procedures for cardiac blood vessels, cerebral blood vessels, abdominal blood vessels and lower extremities.

Recall number
Z-0331-2012
Recalling firm
Toshiba American Medical Systems Inc
Product
Model CAT-850B Catheterization Table, a component of the Toshiba INFX-8000 series X-Ray Interventional System. These systems are single or dual plane systems that employ x-rays to obtain fluoroscopy or radiography images of the body. The system is comprised of a computer system, support arms (that contain the tube and solid state detector), x-ray generator and a patient table. This device is a digital x-ray system configured for use as a diagnostic x-ray angiography system. This x-ray angiography system is used for diagnostic and interventional procedures for cardiac blood vessels, cerebral blood vessels, abdominal blood vessels and lower extremities.
Status
Terminated
Initiated
2011-10-03
Posted
2011-12-01
Terminated
2013-03-18
Reason
The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The result of the tabletop rotating unexpectedly is the patient being tranferred may fall from the tabletop.
Root cause
Device Design
Action
Toshiba America Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 3, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers were instructed to follow the user instructions by positioning the tabletop at the foot-end position, and ensuring sufficient assistance is available, when transferring a patient from the gurney to the CAT-850B catheterization table. Customers were also instructed to follow the instructions for transferring a patient to the CAT-850B that are included in the operation manual. A Customer Reply Form was included with the letter for customers to complete and return via fax to (877) 349-3054. The customer reply form can also be sent via e-mail to raffairs@tams.com. Questions regarding the recall letter can contact the Director of Regulatory Affairs at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
Quantity
300 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Last Four Digits of Lot #2408, 2064, 2851, 2076, 2081, 2804, 2109, 2100, 2519, 2521, 2008, 3063, 2537, 2350, 2382, 2834, 3030, 3043, 3250, 3194, 2105, 2938, 2063, 2148, 2187, 2313, 2324, 2330, 2331, 2424, 2284, 2692, 2067, 2416, 2802, 3294, 2846, 2852, 2901, 3011, 3075, 3167, 2590, 2591, 2462, 3044, 2305, 3155, 3066, 3102, 3290, 3188, 2176, 2781, 3293, 2117, 3308, 3309, 2862, 2321, 2225, 2235, 2238, 3041, 7003, 7004, 2502, 2599, 3317, 2246, 2272, 2217, 3330, 3048, 2954, 3341, 2391, 2835, 2557, 2303, 2294, 2440, 2348, 2597, 3003, 2369, 2500, 2070, 2613, 2299, 2301, 2216, 3146, 3166, 3110, 2262, 2040, 2675, 3064, 2097, 2145, 3203, 2870, 2306, 2750, 2776, 2691, 2410, 7005, 2930, 2494, 2052, 2244, 2245, 2457, 3345, 2683, 2152, 2995, 2970, 2816, 2763, 2593, 2708, 2674, 2744, 2547, 2966, 2967, 2730, 2836, 2843, 2844, 2845, 2643, 2751, 2261, 2373, 2697, 2559, 2405, 2565, 3189, 2121, 3134, 2821, 3195, 2157, 2158, 2206, 3193, 3285, 3350, 2841, 2401, 2503, 2863, 2534, 2140, 2530, 3355, 3262, 3307, 2581, 2736, 2982, 2418, 2280, 2281, 3149, 2353, 2456, 2346, 2349, 2091, 2343, 3213, 2815, 2857, 2909, 2973, 3049, 3047, 2544, 2688, 3060, 2315, 2308, 2676, 2961, 3176, 2760, 3187, 2267, 2504, 2908, 2738, 2007, 3156, 3373, 3380, 2354, 2524, 2190, 2250, 3173, 3197, 3349, 2470, 2178, 2174, 2785, 3022, 3371, 3108, 2394, 3352, 2383, 2824, 2574, 2659, 2789, 2782, 2788, 2498, 2614, 2975, 2398, 3150, 2620, 2752, 3319, 2445, 2501, 2837, 2737, 2801, 2194, 2213, 2513, 2558, 2957, 2984, 2204, 3123, 2395, 2478, 2595, 2921, 3325, 2399, 3074, 3087, 3181, 2900, 2538, 3314, 2082, 3236, 2814, 2095, 3200, 3055, 3056, 2022, 3178, 2310, 3093, 2198, 2096, 2268, 3305, 2952, 2628, 2994, 2427, 2655, 2989, 2374, 2447, 3378, 3198, 3346, 2362, 2455, 2721, 3261, 2423, 2439, 2569, 2682, 3177, 3233, 2104, 2871, 2876, 2073, 2316, 2128, 3336, 3427.
510(k) numbers
["K081582","K081621","K081624","K082830"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register