Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1 Pressure Monitoring; Catalog number 802112.
- Recall number
- Z-0331-2007
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Product
- Terumo Advanced Perfusion System 1 Pressure Monitoring; Catalog number 802112.
- Status
- Terminated
- Initiated
- 2004-10-12
- Posted
- 2006-12-30
- Terminated
- 2007-02-01
- Reason
- Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
- Root cause
- Other
- Action
- U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
- Quantity
- 4,740 for all affected products
- Distribution
- Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
- Lot, serial or model codes
- Serial numbers 00017 through 00032, 00036, 00043, 00045 through 00057, 00059 through 00062, 00065 through 00071, 00073 through 00079, 00081 through 00099, 00102 through 00122, 00125 through 000251, 00253 through 00292, 00296 through 00309, 00311 through 00349, 00352 through 00479, 00484 and 00489.
- 510(k) numbers
- ["K022947"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28