FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography
- Recall number
- Z-0330-2018
- Recalling firm
- Philips Electronics North America Corporation
- Product
- CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography
- Status
- Terminated
- Initiated
- 2017-10-19
- Posted
- not on file
- Terminated
- 2018-09-20
- Reason
- The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.
- Root cause
- Nonconforming Material/Component
- Action
- Philips Medical System sent an Customer Notification letter dated October 19, 2017, and plans to have a Philips Field Service Engineer go to site to perform the field change order (FCO70900041) including: 1. Check for the presence of the PE cable and, if missing, install it. 2 .Perform a modified screw fixation. They expect to complete action by May 2018. For further questions, please call (978) 659-3000.
- Quantity
- 10
- Distribution
- Worldwide Distribution - US Distribution and to the countries of : France, Germany, Luxembourg and New Zealand.
- Lot, serial or model codes
- All CombiDiagnost R90 UDI 709030
- 510(k) numbers
- ["K163210"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28