Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems, Inc. Oncology Systems: Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i; Ref: BT-00264, Model 2-H63; Product is distributed by Varian Medical Systems Inc., Palo Alto, CA and Manufactured by Varian Medical Systems Haan GmbH, Bergische Strasse 16, D-42781 Haan, Germany. The Varian Medical System GammaMed Afterloader system is a remotely controlled intelligent robotic device intended to be used to deliver high dose-rate radiation therapy, in close proximity to and from within diseased tissues. Palliative care and treatments with curative intent to both benign and non-benign disease using interstitial, intracavitary, intraluminal, intra and inter-operative brachytherapy or combinations of any these treatment types are possible using this device.

Recall number
Z-0330-2012
Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Product
Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i; Ref: BT-00264, Model 2-H63; Product is distributed by Varian Medical Systems Inc., Palo Alto, CA and Manufactured by Varian Medical Systems Haan GmbH, Bergische Strasse 16, D-42781 Haan, Germany. The Varian Medical System GammaMed Afterloader system is a remotely controlled intelligent robotic device intended to be used to deliver high dose-rate radiation therapy, in close proximity to and from within diseased tissues. Palliative care and treatments with curative intent to both benign and non-benign disease using interstitial, intracavitary, intraluminal, intra and inter-operative brachytherapy or combinations of any these treatment types are possible using this device.
Status
Terminated
Initiated
2011-10-06
Posted
2011-12-01
Terminated
2012-10-22
Reason
The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.
Root cause
Device Design
Action
Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION LETTER/FIELD SAFETY NOTICE" letter dated October 6, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Contact the Varian Help Desk at 1-888-827-4265 for questions regarding this notice.
Quantity
not on file
Distribution
Worldwide Distribution--USA (nationwide).
Lot, serial or model codes
CODES:   H630164 ,H630703 ,H630727 ,H630756 ,H630787 ,H630810 ,H630829 ,H630850 ,H630868 ,H630888 ,H630165 ,H630705 ,H630728 ,H630757 ,H630789 ,H630811 ,H630832 ,H630851 ,H630869 ,H630889 ,H630168 ,H630706 ,H630729 ,H630758 ,H630790 ,H630812 ,H630833 ,H630852 ,H630870 ,H630891 ,H630170 ,H630711 ,H630732 ,H630760 ,H630792 ,H630813 ,H630834 ,H630854 ,H630871 ,H630892 ,H630205 ,H630712 ,H630733 ,H630764 ,H630793 ,H630814 ,H630835 ,H630855 ,H630872 ,H630895 ,H630206 ,H630713 ,H630736 ,H630769 ,H630794 ,H630815 ,H630837 ,H630857 ,H630877 ,H630896 ,H630210 ,H630714 ,H630737 ,H630771 ,H630795 ,H630817 ,H630839 ,H630858 ,H630878 ,H630897 ,H630211 ,H630716 ,H630739 ,H630774 ,H630798 ,H630818 ,H630841 ,H630859 ,H630879 ,H630898 ,H630214 ,H630718 ,H630741 ,H630776 ,H630801 ,H630820 ,H630842 ,H630860 ,H630880 ,H630901 ,H630215 ,H630719 ,H630744 ,H630777 ,H630802 ,H630822 ,H630843 ,H630863 ,H630881 ,H630902 ,H630216 ,H630720 ,H630747 ,H630778 ,H630804 ,H630825 ,H630845 ,H630864 ,H630884 ,H630903 ,H630219 ,H630721 ,H630749 ,H630781 ,H630807 ,H630826 ,H630847 ,H630866 ,H630885 ,H630904 ,H630701 ,H630722 ,H630753 ,H630784 ,H630809 ,H630828 ,H630848 ,H630867 ,H630886
510(k) numbers
["K891131"]
PMA numbers
not on file
Product code
JAQ
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register