Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems, Inc.: VariSource iX, Software Version 1.1, Part Number AL20020000 VariSource iX Afterloader. The VariSource iX Series High Dose Rate Afterloader system is a computer controlled remote electra/mechanical system used for medical purposes, for placing a NiTiNol wire incorporating an irradiated iridium seed internally or close by, a malignant tumor or tumor bed in a practice known as brachytherapy. The device has up to 20 channels for treatment.

Recall number
Z-0325-2012
Recalling firm
Varian Medical Systems, Inc.
Product
VariSource iX, Software Version 1.1, Part Number AL20020000 VariSource iX Afterloader. The VariSource iX Series High Dose Rate Afterloader system is a computer controlled remote electra/mechanical system used for medical purposes, for placing a NiTiNol wire incorporating an irradiated iridium seed internally or close by, a malignant tumor or tumor bed in a practice known as brachytherapy. The device has up to 20 channels for treatment.
Status
Terminated
Initiated
2010-03-01
Posted
2011-12-01
Terminated
2013-05-30
Reason
VariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment.
Root cause
Software design
Action
Varian Medical Systems sent a "MEDICAL DEVICE RECALL" letter dated April 1, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were asked to conduct a simple test to establish if their machine is affected by the problem. Device users were requested to complete an attached receipt verification form and return it to Varian BrachyTherapy via fax or e-mail. Call the Contact Service Dispatch number at 1-800-864-1672 for questions regarding this detail.
Quantity
72 worldwide/50 in U.S.
Distribution
Worldwide Distribution-USA (nationwide)
Lot, serial or model codes
Serial numbers: 212, 215, 222, 224, 239, 246, 271, 274, 281, 293, 308, 322, 333, 343, 346, 388, 394, 401, 403, 406, 411, 431, 436, 437, 438, 441, 444, 446, 450, 456, 462, 500, 501, 505, 509, 510, 511, 512, 513, 514, 515, 516, 518, 519, 521, 522, 527, 530, 532, 535, 536, 538, 539, 540, 541, 542, 544, 545, 546, 548, 551. Part Number: AL20020000 (VariSource iX Afterloader
510(k) numbers
["K071467"]
PMA numbers
not on file
Product code
JAQ
Firm FEI
3000206172
Firm city
Charlottesville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register