Skip to the content

FDA Medical Device Recalls (openFDA)

Diamedix Corporation: The Is-Syphilis TREP-CHEK (TM) Test Kit is a confirmatory in vitro enzyme immunoassay for the qualitative detection of Treponema pallidum IgG antibodies in human serum or plasma. The reagents can be used either manually or in conjunction with the MAGO Plus or APTUS Automated EIA Processors. The FDA has not cleared this product for use in screening blood or plasma donors. The kit is labeled with an expiration date of Mar 04 and requires storage at 2-8 degree Centigrade.The non conforming kit component is the Conjugate that consists of a glass bottle with red cap containing 30 ml goat anti-human IgG labeled with horseradish peroxidase.

Recall number
Z-0325-05
Recalling firm
Diamedix Corporation
Product
The Is-Syphilis TREP-CHEK (TM) Test Kit is a confirmatory in vitro enzyme immunoassay for the qualitative detection of Treponema pallidum IgG antibodies in human serum or plasma. The reagents can be used either manually or in conjunction with the MAGO Plus or APTUS Automated EIA Processors. The FDA has not cleared this product for use in screening blood or plasma donors. The kit is labeled with an expiration date of Mar 04 and requires storage at 2-8 degree Centigrade.The non conforming kit component is the Conjugate that consists of a glass bottle with red cap containing 30 ml goat anti-human IgG labeled with horseradish peroxidase.
Status
Terminated
Initiated
2003-09-11
Posted
2004-12-17
Terminated
2009-10-05
Reason
Several customers complained that either the negative control index value exceeded the upper control limit or there was very low activity of the cut off calibrator causing patient sample test runs to be called invalid.
Root cause
Other
Action
Recall letter dated 10/8/03 was sent, via FEDEX, with replacement conjugate bottles to those accounts reporting remaining stock.The letter requested that the original conjugate bottle, lot number 42023 be discarded and replaced with new conjugate lot #71323z. The letter also included a new inner box label that had the lot number information for the new conjugate bottle.
Quantity
490 kits
Distribution
Distributed to 54 laboratory accounts throughout the US.
Lot, serial or model codes
The kit lot number is 42003. The conjugate lot number is 42023.
510(k) numbers
["K001552"]
PMA numbers
not on file
Product code
LIP
Firm FEI
1044713
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register