FDA Medical Device Recalls (openFDA)
Pega Medical Inc.: SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. Orthopedic.
- Recall number
- Z-0324-2016
- Recalling firm
- Pega Medical Inc.
- Product
- SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. Orthopedic.
- Status
- Terminated
- Initiated
- 2015-10-20
- Posted
- 2015-11-20
- Terminated
- 2016-08-16
- Reason
- SCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.
- Root cause
- Unknown/Undetermined by firm
- Action
- The firm sent the Urgent Field Safety Notice-Recalls to US consignees via email. Consignees are advised to return affected devices back to the firm. Customer with questions can contact Enrique Garcia at 450-688-5144 ext. 242 or send email to egarcia@pegamedical.com
- Quantity
- 4 units distributed in the US
- Distribution
- Distributed in the states of FL and TN.
- Lot, serial or model codes
- Part Number: SCF-MLD265; LOT#: 130703-01 Part Number: SCF-MLD273; LOT#: 130703-02
- 510(k) numbers
- ["K131591"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 3000327445
- Firm city
- Laval
- Firm state
- not on file
- Firm country
- Canada
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28