Synthes (USA) Products LLC: 2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, particularly in osteopenic bone.

Recall number
Z-0323-2016
Recalling firm
Synthes (USA) Products LLC
Product
2.4MM locking screw SLF-TPNG with Stardrive Recess 18 mm. Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, particularly in osteopenic bone.
Status
Terminated
Initiated
2015-11-02
Posted
not on file
Terminated
2016-07-15
Reason
It was discovered that the above part number and lot of 2.4mm Locking Screws may contain the incorrect package insert (GP2613, VA-LCP Distal Radius System). The correct package insert for the above part is GP2615 (Modular Mini Fragment LCP System).
Root cause
Process control
Action
Synthes sent an "URGENT FIELD SAFETY NOTIFICATION" letter dated November 2, 2015. Customers were asked to take the following actions: -Review the correct package insert (GP2615, attached). -If found, discard incorrect package insert associated with the locking screw noted in the table above. -Forward this Field Safety Notification to anyone in your facility that needs to be informed. -If the package insert has been forwarded to another facility, contact that facility. -Complete the attached Verification Section (page 3 of this notification). Please include your name, title, address, telephone number and signature in the spaces provided. ---Send a copy of the completed Verification Section by: -Fax: 877-907-7514 or Scan/email: Synthes7766@stericycle.com -If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page three (3) of this notification. -Maintain a copy of this Notification.
Quantity
229
Distribution
Distributed in the states of IA, OH, CO, FL, TX, MA, NY, OR, SC, VA, WV, TN, MO, PA, CT, GA, WA, CA, AL, , IN, MI, NJ, VT, ID, KS, MN, and MD, and the country of Canada.
Lot, serial or model codes
Part Number Lot Number Product Description 212.818 7856527 Synthes 2.4mm Locking Screw SLF-TPNG with Stardrive Recess 18mm
510(k) numbers
["K012114","K040588","K050110","K063049","K102694","K103243"]
PMA numbers
not on file
Product code
HWC
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register