Cordis Corporation: Product is packaged and labeled as: ''Cordis PRECISE Nitinol Stent Transhepatic Billiary System'', Catalog Nos. N820SB, N830SB, N840SB, N920SB, N930SB, N940SB, N1020SB, N1030SB, N1040SB only. This recall also affects Cordis PRECISE Nitinol Stent System, not distributed in the United States, Catalog numbers, N920SC, N930SC, N940SC, N1020SC, N1030SC and N1040SC only.
- Recall number
- Z-0323-03
- Recalling firm
- Cordis Corporation
- Product
- Product is packaged and labeled as: ''Cordis PRECISE Nitinol Stent Transhepatic Billiary System'', Catalog Nos. N820SB, N830SB, N840SB, N920SB, N930SB, N940SB, N1020SB, N1030SB, N1040SB only. This recall also affects Cordis PRECISE Nitinol Stent System, not distributed in the United States, Catalog numbers, N920SC, N930SC, N940SC, N1020SC, N1030SC and N1040SC only.
- Status
- Terminated
- Initiated
- 2002-11-07
- Posted
- 2002-12-05
- Terminated
- 2008-11-25
- Reason
- Field complaints of marker band spacing being out of spec.
- Root cause
- Other
- Action
- The firm sent, via First Class Mail, a Recall notification Letter with Response Card on 11/7/02. The letter asks for response from the user accounts and for return of product. The letter includes a UPS label for return of product. Non-responders will be recontacted as part of the firm''s effectiveness checks.
- Quantity
- 10,468
- Distribution
- Product was distributed to 620 hospital accounts nationwide. The firm did not provide information on international accounts. This info was requested from firm and RES will be updated when available.
- Lot, serial or model codes
- All lot numbers of the indicated Product/Catalog numbers.
- 510(k) numbers
- ["K012993"]
- PMA numbers
- not on file
- Product code
- FGE
- Firm FEI
- not on file
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28