Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc: Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5

Recall number
Z-0322-2024
Recalling firm
Medtronic Inc
Product
Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5
Status
Open, Classified
Initiated
2023-10-12
Posted
2023-11-17
Terminated
not on file
Reason
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
Root cause
Process control
Action
Medtronic issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION on 10/04/2023 via UPS. The notice explained the issue, potential risk, and requested the return of the device. For questions, consignees were directed to contact their Medtronic Field Sales Representative.
Quantity
48 units
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.
Lot, serial or model codes
UDI/DI 08033477090696:
510(k) numbers
["K092177"]
PMA numbers
not on file
Product code
NTE
Firm FEI
3001504994
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register