FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Somatom Sensation 64 X-Ray System, Model number 8377520. This is a computed tomography x-ray system which is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles.
- Recall number
- Z-0321-2009
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Somatom Sensation 64 X-Ray System, Model number 8377520. This is a computed tomography x-ray system which is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles.
- Status
- Terminated
- Initiated
- 2008-09-03
- Posted
- 2009-01-05
- Terminated
- 2009-09-23
- Reason
- software error - does not allow protocol to properly synchronize with ECG triggers
- Root cause
- Software design
- Action
- The recalling firm issued a Customer Safety Advisory Notice dated 9/3/08 to affected customers via Update Instructions CT045/08/S. This letter informs customers of the potential issued and provides instruction to avoid its occurrence. For additional information, contact Siemens Medical Solutions USA, Inc. at 1-610-448-4500.
- Quantity
- 9 units
- Distribution
- Nationwide
- Lot, serial or model codes
- Serial numbers 57084, 57193, 57213, 57261, 57262, 57270, 57272, 57285, and 57298
- 510(k) numbers
- ["K040665"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28