FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application

Recall number
Z-0319-2021
Recalling firm
Boston Scientific Corporation
Product
Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application
Status
Terminated
Initiated
2020-09-22
Posted
not on file
Terminated
2024-09-04
Reason
There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.
Root cause
Software design
Action
Boston Scientific sales representatives will hand-deliver a removal letter to all customer consignees with one of the eight affected Model 3300 LATITUDE programmer. BSC sales representatives will maintain control of these devices. In order to continue to provide the benefits that this software support application provides for those patients implanted with an ALTRUA/INSIGNIA I pacemaker, BSC sales representatives will continue to use Model 3300 LATITUDE programmers installed with the Model 3892 v1.04 ALTRUA / INSIGNIA I / NEXUS I Software Support Application in the field. The risks associated with this unintended behavior can be avoided when specific remedial actions are adhered to. These remedial actions will be adhered to until a software correction is approved and installed on these programmers. BSC will confirm via email from sales representatives that 100% of the removal letters were delivered. BSC will post the removal letter on the BSC website within the Product Performance Resource Center located @ www.bostonscientific.com/en-US/pprc/productadvisories.html.
Quantity
8 devices
Distribution
Distributed in US - AR, FL, IL, and TX
Lot, serial or model codes
Model 3892 v1.04 is installed on Model 3300 LATITUDE Programming Systems with the following serial numbers: 009494 012689 010356 008206 008124 007983 004813 010751
510(k) numbers
not on file
PMA numbers
["P910077S178"]
Product code
LWS
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.