Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Stereotactic Circular Collimator Product Usage: This device is intended to hold a patient's head in a fixed position and to localize and center the output of a linear accelerator (UNAC) to allow radiotherapy of brain tumors and other types of cerebral lesions.

Recall number
Z-0319-2012
Recalling firm
Elekta, Inc.
Product
Stereotactic Circular Collimator Product Usage: This device is intended to hold a patient's head in a fixed position and to localize and center the output of a linear accelerator (UNAC) to allow radiotherapy of brain tumors and other types of cerebral lesions.
Status
Terminated
Initiated
2011-03-25
Posted
2011-11-30
Terminated
2012-10-02
Reason
Recent newspaper articles outlined improper use of SRS Cone Collimator accessories that injured patients on Brainlab and Varian systems.
Root cause
Under Investigation by firm
Action
Elekta sent an IMPORTANT NOTICE letter dated March 25, 2011 to all affected customers. The letter identified the affected product, the problem, key safety procedures and information to users for safe operation of the Stereotactic Circular Collimator. Elekta will issue a Mandatory Action Field Change Order which will include an updated Circular Collimator Instructions for Use - 102022201 and new labels. The notice instructs Users to file this document in the Important Notice section of the appropriate User Manual. Also, Elekta is offering some free WebEx sessions which will include the procedures and recommended operation of these types of accessories. Customers are instructed to register at www.elekta.com/SafeAccessories.
Quantity
89 devices
Distribution
Nationwide Distribution (USA) - including the states of: AZ, CA, FL, GA, MO, NJ, NM, NC, OR, TN, and WV
Lot, serial or model codes
848, 849, 850, 851, 852, 853, 1032, 1033, 1034, 928, 929, 930, 931, 932, 801, 802, 803, 804, 805, 980, 981, 982, 983, 984, 889, 890, 891, 892, 893, 894, 908, 909, 910, 911, 912, 833, 834, 835, 836, 837, 843, 844, 845, 846, 847, 1017, 1018, 1019, 1020, 1021, 903, 904, 905, 906, 907, 883, 884, 885, 886, 933, 935, 936, 937, 940, 974, 975, 976, 977, 978, 941, 942, 943, 944, 945, 946, 947, 948, 949, 950, 1012, 1013, 1014, 1015, 1016, 1027, 1028, 1029, 1030, 1031
510(k) numbers
["K011255"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register