GE Healthcare: GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, 105468-1-Centricity Universal Viewer Zero Footprint / ZFP-02254840, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090, JP-340111-Centricity Universal Viewer Zero Footprint / ZFP-01275379, JP-275353-Centricity Universal Viewer Zero Footprint / ZFP-02042636; 6) 2089507-112, 3922-25-Centricity Universal Viewer Zero Footprint / ZFP-01290685, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 1503586-Centricity Universal Viewer Zero Footprint / ZFP-21016603, RU0994-Centricity Universal Viewer Zero Footprint / ZFP-02331255; 7) 2089507-114, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-02360229, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225897, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225898, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225904, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01371500, M51725-Centricity Universal Viewer Zero Footprint / ZFP-01300873, M1781854-Centricity Universal Viewer Zero Footprint / ZFP-01553772, 609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, CMC-Centricity Universal Viewer Zero Footprint / ZFP-01820241, JP-100014-Centricity Universal Viewer Zero Footprint / ZFP-01890709, KW1102-Centricity Universal Viewer Zero Footprint / ZFP-02191178, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02327073, QA1114-Centricity Universal Viewer Zero Footprint / ZFP-02306397, SAU1433-Centricity Universal Viewer Zero Footprint / ZFP-01457278, TTSH-CVIS-Centricity Universal Viewer Zero Footprint / ZFP-02041523, NE4651-Centricity Universal Viewer Zero Footprint / ZFP-00038599, NE6703-Centricity Universal Viewer Zero Footprint / ZFP-00038552, NE6731-Centricity Universal Viewer Zero Footprint / ZFP-00038461, NE3006-Centricity Universal Viewer Zero Footprint / ZFP-00038362, NE2196-Centricity Universal Viewer Zero Footprint / ZFP-00038588, NPV738-Centricity Universal Viewer Zero Footprint / ZFP-00038637, NE2031-Centricity Universal Viewer Zero Footprint / ZFP-00038578, NE7405-Centricity Universal Viewer Zero Footprint / ZFP-00038372, NE2162-Centricity Universal Viewer Zero Footprint / ZFP-00038563, NE4878-

Recall number
Z-0318-2025
Recalling firm
GE Healthcare
Product
GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, 105468-1-Centricity Universal Viewer Zero Footprint / ZFP-02254840, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090, JP-340111-Centricity Universal Viewer Zero Footprint / ZFP-01275379, JP-275353-Centricity Universal Viewer Zero Footprint / ZFP-02042636; 6) 2089507-112, 3922-25-Centricity Universal Viewer Zero Footprint / ZFP-01290685, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 1503586-Centricity Universal Viewer Zero Footprint / ZFP-21016603, RU0994-Centricity Universal Viewer Zero Footprint / ZFP-02331255; 7) 2089507-114, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-02360229, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225897, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225898, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225904, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01371500, M51725-Centricity Universal Viewer Zero Footprint / ZFP-01300873, M1781854-Centricity Universal Viewer Zero Footprint / ZFP-01553772, 609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, CMC-Centricity Universal Viewer Zero Footprint / ZFP-01820241, JP-100014-Centricity Universal Viewer Zero Footprint / ZFP-01890709, KW1102-Centricity Universal Viewer Zero Footprint / ZFP-02191178, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02327073, QA1114-Centricity Universal Viewer Zero Footprint / ZFP-02306397, SAU1433-Centricity Universal Viewer Zero Footprint / ZFP-01457278, TTSH-CVIS-Centricity Universal Viewer Zero Footprint / ZFP-02041523, NE4651-Centricity Universal Viewer Zero Footprint / ZFP-00038599, NE6703-Centricity Universal Viewer Zero Footprint / ZFP-00038552, NE6731-Centricity Universal Viewer Zero Footprint / ZFP-00038461, NE3006-Centricity Universal Viewer Zero Footprint / ZFP-00038362, NE2196-Centricity Universal Viewer Zero Footprint / ZFP-00038588, NPV738-Centricity Universal Viewer Zero Footprint / ZFP-00038637, NE2031-Centricity Universal Viewer Zero Footprint / ZFP-00038578, NE7405-Centricity Universal Viewer Zero Footprint / ZFP-00038372, NE2162-Centricity Universal Viewer Zero Footprint / ZFP-00038563, NE4878-
Status
Open, Classified
Initiated
2024-10-03
Posted
2024-11-07
Terminated
not on file
Reason
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
Root cause
Software Manufacturing/Software Deployment
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to it consignees on 10/03/2024 via letter issued through a traceable means. The notice explained the issue and provided the following actions be to be implemented until the devices can be corrected: Option 1: Restrict access to these systems from the internet. or Option 2: Implement additional safeguards such as a VPN. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative. GE HealthCare will correct all affected products at no cost. A GE HealthCare representative will make contact to arrange for the correction.
Quantity
120 units
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.
Lot, serial or model codes
UDI/DI 00840682102988 for all models: 1) 2089507-083, Installed Product ID Numbers: GDZH-JINWAN-Centricity Universal Viewer Zero Footprint / ZFP-02171477, 103378-Centricity Universal Viewer Zero Footprint / ZFP-20039058, 1664823-Centricity Universal Viewer Zero Footprint / ZFP-20080654, 493356-Centricity Universal Viewer Zero Footprint / ZFP-20085691, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291850, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291877, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02292580, 81077-Centricity Universal Viewer Zero Footprint / ZFP-20680055, IL1469-Centricity Universal Viewer Zero Footprint / ZFP-02156638, JP-110145-Centricity Universal Viewer Zero Footprint / ZFP-01275393, 1333629-Centricity Universal Viewer Zero Footprint / ZFP-01300332; 2) 2089507-093, Installed Product ID Numbers: 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, Installed Product ID Numbers: JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, Installed Product ID Numbers: 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, Installed Product ID Numbers: 6) 2089507-112, Installed Product ID Numbers: 7) 2089507-114, Installed Product ID Numbers: 8) 2089507-115, Installed Product ID Numbers: 9) 2089507-116, Installed Product ID Numbers: 10) 2089507-117, Installed Product ID Numbers: 11) 2089507-118, Installed Product ID Numbers: 12) 2089507-119, Installed Product ID Numbers:
510(k) numbers
["K131977"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3004526608
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register