Skip to the content

FDA Medical Device Recalls (openFDA)

Atrium Medical Corporation: Express Dry Seal Chest Drains (Express): 4000-100N DRAIN, EXPRESS, SINGLE; 4050-100N DRAIN, EXPRESS, BRU

Recall number
Z-0318-2024
Recalling firm
Atrium Medical Corporation
Product
Express Dry Seal Chest Drains (Express): 4000-100N DRAIN, EXPRESS, SINGLE; 4050-100N DRAIN, EXPRESS, BRU
Status
Open, Classified
Initiated
2023-09-18
Posted
2023-11-16
Terminated
not on file
Reason
Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
Root cause
Error in labeling
Action
UPDATE: April 2024. Getinge issued letter Getinge Thoracic Drainage Labeling Updates: New Instructions for Use Precaution for Oasis and Express Chest Drains to existing customers April 17, 2024 (US) and April 23, 2024 (OUS). As a result, inventory will be a mix of the current IFU and new IFU with the updated precaution statement until the current IFU-containing products are depleted. Reminder: In September 2023, Getinge initiated Medical Device Correction 3011175548-08/18/2023-001-C which requested your facility to post a copy of the Notice on Page 4 of the notification in all inventory locations where the devices are stored. Since the updated IFU now contains this new Precaution, the Page 4 Notice may be removed from all inventory locations once all affected product with expiry labeled March 7th, 2024 and earlier has been depleted. Please contact your Getinge Inside Sales Representative should you need an electronic copy of the new Instructions for Use. Atrium/Getinge issued URGENT MEDICAL DEVICE Correction letter on September 18, 2023 via FedEx 2 day delivery on or before, September 20, 2023. Letter states reason for recall, health risk and action to take: facility can continue use of the device. No devices need to be returned. " Please ensure that all Atrium Ocean, Oasis, or Express Single Collection Chest Drain users at your facility are aware of this Safety Notice and post a copy of the Notice on Page 4 in all inventory locations within your facility where the devices are stored Notification of the release of the updated IFU containing the new Precaution will be communicated to all customers upon release, including reminder the Notice on Page 4 may be removed upon the facility s receipt of product with the updated IFU. " Please forward this information to all current and potential Atrium Ocean, Oasis, or Express Single Collection Chest Drain users within your hospital / facility. " If you are a distributor who has shipped any affected products to custom
Quantity
48339 cases: 48148 cases US; 193 cases OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Brazil, Brunei Darussalam, Canada, Chile, China, Czech Republic Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Korea, Republic of Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Norway, Pakistan, Peru, Poland, Portugal, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan Province of China, Thailand Tunisia, Ukraine, United Arab Emirates, United Kingdom, Uruguay.
Lot, serial or model codes
UDI-DI: (1)4000-100N DRAIN, EXPRESS, SINGLE 20650862115134; (2) 4050-100N DRAIN, EXPRESS, BRU 20650862115141 All lots within labeled product expiry Manufacturing Dates: Any product manufactured on and after July 21, 2020 Distribution Dates: Any product shipped on and after August 6, 2020
510(k) numbers
["K043140","K201305"]
PMA numbers
not on file
Product code
CAC
Firm FEI
3011175548
Firm city
Merrimack
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register