FDA Medical Device Recalls (openFDA)

Community Products, LLC: Toddler Chair Hip Strap (Belt). The hip belt is made with nylon webbing and a plastic buckle. 510(k) exempt, Medical Device Listing # R061877. The product is used as a hip strap chair for children.

Recall number
Z-0318-2009
Recalling firm
Community Products, LLC
Product
Toddler Chair Hip Strap (Belt). The hip belt is made with nylon webbing and a plastic buckle. 510(k) exempt, Medical Device Listing # R061877. The product is used as a hip strap chair for children.
Status
Terminated
Initiated
2008-09-12
Posted
2008-12-05
Terminated
2010-03-15
Reason
Some hip straps were assembled incorrectly by threading the buckle on the strap upside down. If a child were left unattended in spite of the warnings on the product and in the manual, a defective hip strap could loosen, allowing the child to slip down in the chair and this could create a choking hazard.
Root cause
Process control
Action
On 9/12/08, Rifton Equipment began notifying the customers by phone using a phone script. This phone call notified the customers of the recall and also served as an effectiveness check that they were notified. The customers were instructed to remove the product from use until the replacement kit is received and the old strap is returned. The replacement kit contains instructions with ID codes and a place for the customer to verify that they have made the replacement as instructed. It also contains the required number of new hip straps. A verification check was sent with the replacement straps, which the customer will sign and return to the firm verifying that they have replaced the straps for the affected serial numbers. Rifton Equipment is providing the customer with return shipping labels which were sent by UPS 2nd day air. The replacement straps use black thread instead of tan thread for the sewing on the ends of the straps so that the firm can be sure that the returned straps are not the new replacement straps. For questions or additional information, contact Rifton Equipment at 1-800-571-8198.
Quantity
Potentially 147 units may be affected.
Distribution
Worldwide Distribution including USA and countries of Spain and Great Britain.
Lot, serial or model codes
Serial Numbers: E77020080701AE, E77020080701AF, E78020080703AB, E78020080703AC, E77020080701AC, E78020080703AD, E77020080701AD,  E77020080701AA, E77020080701AB, E76020080704AA, E78020080704AA,  E77020080704AE, E77020080704AA, E77020080704AB, E77020080704AC, E78020080703AA, E78020080705AB, E78020080705AA, E76020080709AB, E78020080709AA, E76020080709AA, E78020080709AB, E77020080709AA,  E77020080709AB, E77020080709AC, E77020080709AD, E78020080712AD, E77020080711AA, E77020080712AA, E77020080709AE, E78020080711AB, E78020080712AC, E76020080712AA, E78020080711AA, E78020080712AA, E78020080712AB, E77020080704AD, E78020080715AD, E76020080716AC, E77020080716AF, E78020080715AE, E78020080715AF, E78020080715AG, E78020080716AA, E76020080716AB, E77020080716AA, E77020080716AB, E78020080716AC, E76020080716AA, E78020080716AB, E77020080714AA, E77020080714AB, E77020080714AC, E77020080714AD, E77020080714AE, E78020080707AA, E78020080715AA, E78020080715AB, E78020080715AC,  E78020080719AA, E77020080716AC, E77020080716AD, E76020080722AA, E78020080719AB, E78020080723AA, E77020080723AB, E77020080723AA, E78020080722AA, E77020080722AA, E77020080723AC, E77020080723AD, E77020080716AE, E77020080716AG, E78020080724AD, E78020080724AE,  E78020080724AB, E76020080723AA, E77020080723AE, E78020080724AA, E77020080724AA, E78020080724AC, E78020080723AB, E78020080723AC, E77020080724AB, E77020080725AC, E78020080725AA, E77020080725AA,  E77020080725AB, E78020080728AB, E77020080728AA, E78020080728AA,  E76020080730AA, E77020080730AA, E76020080730AB, E77020080731AB, E78020080802AB, E77020080802AD, E78020080802AA, E78020080802AC, E78020080801AB, E77020080802AB, E77020080802AC, E78020080801AC,  E77020080801AF, E77020080801AA, E78020080716AD, E78020080801AA, E78020080802AD, E77020080801AB, E77020080801AC, E77020080801AD,  E77020080801AE, E77020080804AA, E78020080802AE, E77020080804AC, E78020080804AB, E78020080804AA, E77020080804AB, E77020080806AA, E76020080731AA, E77020080731AA, E77020080802AA, E77020080809AA,  E77020080809AB, E78020080806AF, E77020080806AB, E78020080806AA, E77020080809AC, E78020080809AC, E78020080809AA, E78020080806AB, E78020080806AC, E78020080806AD, E78020080806AE, E78020080809AB,  E76020080809AA, E76020080811AA, E77020080811AA, E78020080812AA, E78020080812AB, E78020080812AC, E78020080812AD, E78020080812AF, E78020080811AA, E78020080812AE, E78020080812AG and E78020080812AH. The code is located underneath the seat of the chair.
510(k) numbers
not on file
PMA numbers
not on file
Product code
INN
Firm FEI
1319061
Firm city
Rifton
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register