FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: Flow Sensor Service Part, Catalog 2096513-001-S
- Recall number
- Z-0317-2021
- Recalling firm
- GE Healthcare, LLC
- Product
- Flow Sensor Service Part, Catalog 2096513-001-S
- Status
- Open, Classified
- Initiated
- 2020-09-17
- Posted
- not on file
- Terminated
- not on file
- Reason
- Flow sensors could have damaged tubes in the form of small punctures or cuts.
- Root cause
- Process control
- Action
- GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
- Quantity
- 2 units
- Distribution
- U.S. states where product was distributed - AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI Foreign countries impacted - Australia, Canada, Czech Republic, Estonia, France, Germany, India, Indonesia, Israel, Italy, Kuwait, Latvia, Macedonia, Madagascar, Malaysia, Nicaragua, Panama, Qatar, Russia, Singapore, Slovenia, Spain, Thailand, Ukraine, United Kingdom
- Lot, serial or model codes
- Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020). This flow sensor is compatible with the following device models: Aespire 7100/11, Aespire 7900, Aespire View, Aestiva 7100, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, Avance CS2, Amingo (OUS), and 9100c NXT (OUS)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28