Skip to the content

FDA Medical Device Recalls (openFDA)

Repro-Med Systems, Inc.: Freedom60 Precision Flow Rate Tubing Sets, REF/Part Numbers F900, F2400, F60, F10, F275, F600, F45, F1200 F180.

Recall number
Z-0317-2018
Recalling firm
Repro-Med Systems, Inc.
Product
Freedom60 Precision Flow Rate Tubing Sets, REF/Part Numbers F900, F2400, F60, F10, F275, F600, F45, F1200 F180.
Status
Terminated
Initiated
2016-03-10
Posted
2018-01-05
Terminated
2018-02-26
Reason
On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.
Root cause
Nonconforming Material/Component
Action
On March 10, 2016 RMS Medical Products distributed Urgent Voluntary Medical Device Corrections and Removal notices to their customers. On March 12, 2016 & on March 18, 2016, RMS Medical Products distributed a 2nd Urgent Voluntary Medical Device Corrections and Removal notices to their customers advising them that affected products list was reduced after a comprehensive risk assessment was performed. Customers are advised to check stock, quarantine, cease distribution and remove product from use. If customers further distributed product, they are also encouraged to inform the affected parties. Customers are encouraged to complete and return the acknowledgement form via fax to: 845-469-5518.
Quantity
not on file
Distribution
Nationwide in US
Lot, serial or model codes
Lot Numbers: 1.053/16, 1.052/16, 1.051/16, 1.050/16, 1.049/16, 1.048/16, 1.047/16, 1.042/16, 1.041/16, 1.040/16, 1.039/16, 1.038/16, 1.037/16, 1.036/16, 1.033/16, & 1.030/16  Expiration: 2021-02
510(k) numbers
["K933652"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1318360
Firm city
Chester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register