Centurion Medical Products Corporation: SafetyGlide Needle, 22GA 1-1/2in, Catalog n. 301701, manufactured by BD, sold in Centurion Convenience Kits. Centurion component code BD17O1NS.
- Recall number
- Z-0315-2018
- Recalling firm
- Centurion Medical Products Corporation
- Product
- SafetyGlide Needle, 22GA 1-1/2in, Catalog n. 301701, manufactured by BD, sold in Centurion Convenience Kits. Centurion component code BD17O1NS.
- Status
- Terminated
- Initiated
- 2017-07-05
- Posted
- not on file
- Terminated
- 2018-03-09
- Reason
- Centurion is recalling specific lots of convenience kits, SafetyGlide Needle, 22GA 1-1/2in, manufactured by BD, sold in Centurion Convenience Kits. distributed between May 2016 and June 2017 due to notice received from BD's recall due to particulate that could be expelled out of the needle if used for injection.
- Root cause
- Material/Component Contamination
- Action
- Centurion Medical Products Corporation sent a notification mailed to the affected customers and distributors via Certified Mail Return Receipt on July 5, 2017
- Quantity
- not on file
- Distribution
- US.
- Lot, serial or model codes
- Material Description Batch Expiration 1 CHT122O CHESTTUBETRAY 2017032350 2018-10 2017032250 2018-06 2017030650 2018-08 2017010550 2018-06 2016121450 2018-06 2016120750 2018-06 2 CHT13OS CHEST TUBE INSERTION TRAY 2017030750 2018-07 3 CHT154O CHEST TUBE INSERTION TRAY 2016112850 2018-06 4 CVI3750 PIGTAIL ACCESSORY BUNDLE 2016121550 2018-09 2016111750 2018-08 5 CV14265 VASCULAR ACCESS INSERT KIT - PEDIATRICS 2017032850 2018-07 6 CV14360 DIALYSIS KIT 2017032250 2018-07 7 PT16O LUMBAR PUNCTURE TRAY 2016120950 2018-03 8 TC7910 TRACHEOTOMY TRAY 2016051950 2018-07 2016050250 2018-07
- 510(k) numbers
- ["K951254"]
- PMA numbers
- not on file
- Product code
- FMI
- Firm FEI
- 3008403546
- Firm city
- Williamston
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28