Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Medfusion, Syringe Infusion Pump, Software Version 6 (V6), Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids.

Recall number
Z-0314-2018
Recalling firm
Smiths Medical ASD Inc.
Product
Medfusion, Syringe Infusion Pump, Software Version 6 (V6), Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids.
Status
Terminated
Initiated
2017-08-22
Posted
not on file
Terminated
2025-04-14
Reason
Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.
Root cause
Component design/selection
Action
The firm,Smiths Medical, sent an, "Urgent Medical Device Recall" Amendment Notice (amending the original August 2017) concerning invalid syringe size alarms on certain Medfusion Syringe Pump Model Series 3500 and 4000. The letter described the product, problem, actions to be to taken, and updated changes, as follows: 1. Locate the affected Medfusion Syringe Pump(s) by referring to the list of affected devices on the attached Urgent Medical Device Recall Notice Response Form. This form provides the specific serial number(s) your organization purchased. Each pump has a unique serial number which can be found on the label located on the bottom of the pump. If you have ten (10) or more affected pumps in your possession, please select one of the following repair options: a. Repair at your Facility: Contact Smiths Medical to schedule a technician to visit your facility and complete the necessary repairs onsite. b. Send to Smiths Medical for Repair: Impacted pumps may be sent to a Smiths Medical Service Center, where repairs will be completed. A prepaid shipping label will be provided. Repairs will be completed within ten (10) business days of receipt. Loaner units are available upon request. c. NEW OPTION - Biomedical Department self-repair at your Facility: Contact Smiths Medical to receive the necessary parts and work instruction to have your biomed team complete the necessary repairs onsite. If you have nine (9) or fewer affected pumps in your possession: a. Send to Smiths Medical for Repair: Impacted pumps may be sent to a Smiths Medical Customer Service Center, where repairs will be completed. A prepaid shipping label will be provided. Repairs will be completed within ten (10) business days of receipt. Loaner units are available upon request. b. NEW OPTION - Biomedical Department Self-Repair at Your Facility: Contact Smiths Medical to receive the necessary parts and work instruction to have your biomed team complete the necessary repairs onsite. To initi
Quantity
9,787 units in total
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Canada, Great Britain, Panama, Germany, India, Philippines, and Oman.
Lot, serial or model codes
Model No. 3500; Reference No. 3500-0600-51
510(k) numbers
["K040899"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register