Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter, Inc.: The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Recall number
Z-0313-2025
Recalling firm
Beckman Coulter, Inc.
Product
The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
Status
Open, Classified
Initiated
2024-09-30
Posted
2024-11-06
Terminated
not on file
Reason
The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.
Root cause
Process design
Action
Beckman Coulter issue and URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2024 via letter (USPS) and email. The notice explained the issue, the impact to the patient, and requested the use of alternative lot of Access Cortisol reagent. In the event the laboratory cannot generate a passing calibration curve use alternative methodology including alternative Beckman Coulter Access Immunoassay Systems. Beckman Coulter will release an updated Access Cortisol APF on the DxI 9000 Access Immunoassay Analyzer that includes the correct calibration curve acceptance parameter at the S1 calibrator level. A Beckman Coulter representative will contact you when the updated APF is available. The notice is to be shared with laboratory staff and forwarded to all parties to whom the the affected product was distributed.
Quantity
149 units
Distribution
Worldwide distribution.
Lot, serial or model codes
UDI/DI 15099590231118, all lots
510(k) numbers
["K223038"]
PMA numbers
not on file
Product code
CGR
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register