Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Recall number
Z-0313-2019
Recalling firm
Becton Dickinson & Co.
Product
BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.
Status
Terminated
Initiated
2018-09-24
Posted
not on file
Terminated
2020-04-10
Reason
A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.
Root cause
Nonconforming Material/Component
Action
Urgent Medical Device Product Recall notification letters dated 9/24/18 were sent to customers.
Quantity
49
Distribution
Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. Countries of Spain, Germany, France, Austria, Denmark, Switzerland, Sweden, Italy, Kuwait, Oman, Saudi Arabia, and Turkey.
Lot, serial or model codes
Serial Numbers: CTI1201-CTI1288
510(k) numbers
not on file
PMA numbers
not on file
Product code
PQA
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.