Skip to the content

FDA Medical Device Recalls (openFDA)

Ameditech Inc: Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5; Immutest 11 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-11OP; Immutest 12 Panel Cup w/Adulteration, Item No. IMCA-12BUP; Immutest 5 panel, Item No. IMCA-5AB-W; Immutest 5 Pnl Drug Cup w/ Adult -waived, Item No. IMCA-5AB-W; Immutest 7 panel, Item No. IMCA-7P; ImmuTest Cup 7-panel w/ Adulteration, Item No. IMCA-7P These Drugs of Abuse Tests are one-step immunoassays for the qualitative detection of multiple drugs and drug metabolites in human urine. The configurations of these Drugs of Abuse Tests consist of a combination of drug strips, including Benzodiazepine (BZO). They are used to obtain a visual, qualitative result and are intended for professional and point of care use.

Recall number
Z-0312-2016
Recalling firm
Ameditech Inc
Product
Immutest 10 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-10M5; Immutest 11 Pnl Drug Screen Cup w/ Adult, Item No. IMCA-11OP; Immutest 12 Panel Cup w/Adulteration, Item No. IMCA-12BUP; Immutest 5 panel, Item No. IMCA-5AB-W; Immutest 5 Pnl Drug Cup w/ Adult -waived, Item No. IMCA-5AB-W; Immutest 7 panel, Item No. IMCA-7P; ImmuTest Cup 7-panel w/ Adulteration, Item No. IMCA-7P These Drugs of Abuse Tests are one-step immunoassays for the qualitative detection of multiple drugs and drug metabolites in human urine. The configurations of these Drugs of Abuse Tests consist of a combination of drug strips, including Benzodiazepine (BZO). They are used to obtain a visual, qualitative result and are intended for professional and point of care use.
Status
Terminated
Initiated
2015-10-20
Posted
not on file
Terminated
2016-11-09
Reason
Ameditech is recalling drug abuse tests because they have shown reduced reactivity.
Root cause
Nonconforming Material/Component
Action
Ameditech Inc. sent an Urgent Medical Device Recall Letter to al affected customers on October 20, 2015, to inform them that Ameditech is recalling affected lots of Ameditech's Drugs of Abuse Tests because they have shown reduced reactivity for up to three of the nineteen Benzodiazepine (BZO) compounds. The letter informs customers of the actions to be taken. Customers were instructed to complete and return the customer verification form to (858) 677-0243 or scan and email to amt.fieldaction@alere.com. For questions regarding this recall call 858-535-1968.
Quantity
1172 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia.
Lot, serial or model codes
144380 144564 150505 150908 151538 142098 144103 150313 150909 141645 144563 150483 151337 142076 142966 144263 151540
510(k) numbers
["K052115","K113046","K113501"]
PMA numbers
not on file
Product code
JXM
Firm FEI
3003789989
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register