Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Roche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.

Recall number
Z-0311-2012
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Roche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Status
Terminated
Initiated
2011-10-18
Posted
2011-11-30
Terminated
2013-02-21
Reason
Internal investigations have revealed a potential printer malfunction on the Cobas c 111 analyzer, where one or more numbers or characters can be missing on the hard copy result printout, or the printer may skip leaving blanks which may result in an incorrect print layout. The missing number of characters can be in any position, therefore it could be in the middle of a result, (e.g. 48.53 mmol in
Root cause
Software design
Action
Roche Diagnostics Operations, Inc sent a Urgent Medical Device Correction dated October 18, 2011, via UPS Ground (receipt required) to all affected customers and one distribution center, with instructions to disable the Auto Print Result feature until new software is available. All printed results will need to be confirmed against the results indicated on the display screen. Questions are to be directed to Roche Diagnostics Technical Support at 1-800 428-2336,
Quantity
135
Distribution
Worldwide Distribution - USA including West Virginia and Texas.
Lot, serial or model codes
Serial Numbers 50865, 50626, 50798, 50731, 50654, 50541, 51023, 50946, 50989, 50634, 50506, 50251, 50887, 50948, 50453, 50688, 51111, 50997, 51122, 50244, 50790, 50475, 50597, 50740, 50656, 50793, 51086, 50999, 51146, 50635, 50744, 50586, 50632, 50846, 1167, 50693, 50787, 50670, 50672, 51128, 50941, 50562, 50788, 50547, 51003, 50624, 50677, 50678, 50814, 51015, 50687, 50680, 50894, 50560, 51004, 50247, 50538, 51118, 50926, 50847, 50841, 51127, 50741, 50764, 50633, 51106, 51116, 50201, 51124, 50612, 51088, 50576, 51016, 50689, 50681, 50424, 51126, 50959, 50878, 50784, 50661, 50761, 50452, 50450, 50290, 50947, 51327, 50630, 50985, 50763, 50319, 50984, 50250, 50549, 50925, 50844, 50660, 50845, 50683, 50735, 51056, 50463, 50775, 50840, 50734, 50655, 50843, 50758, 50852, 50539, 50651, 50653, 50657, 50964, 50563, 50742, 50738, 50727, 50934, 50561, 50803, 51000, 51107, 50686, 50815, 50649, 50470, 50759, 50644, 51053, 50739, 50595, 51087, 50575, and 50679.
510(k) numbers
["K071211"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register