Roche Diagnostics Operations, Inc.: cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.

Recall number
Z-0310-2012
Recalling firm
Roche Diagnostics Operations, Inc.
Product
cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Status
Terminated
Initiated
2011-10-26
Posted
2011-11-30
Terminated
2012-07-09
Reason
The cobas c 311 analyzer software may incorrectly calculate the volume of the ISE Reference Electrolyte solution remaining on the analyzer. In some cases the incorrect results obtained will not be flagged with any data alarms.
Root cause
Software design
Action
October 26, 2011 Urgent Medical Device Corrections were sent to all direct accounts in the USA that have installed the cobas c 311 analyzer and included a faxback form to be completed and faxed to 1-877-766-7452. Consignees were instructed to temporarily replace the ISE Reference Electrolyte bottle based upon the frequency calculated from the maximum number of ISE samples run per day and activate the yellow threshold alarm on the system. Customers with question and concerns are directed to call Roche Diagnostics Technical Support at 1-800-428-2366, available 24-hours, 7 days a week.
Quantity
940 worldwide
Distribution
Worldwide Distribution -- USA including the states of Maryland and Texas.
Lot, serial or model codes
Serial Numbers:1050-16, 1040-08, 1041-06, 0927-16, 0918-05, 1048-02, 0812-12, 1058-09, 0925-20, 1044-19, 1041-05, 1043-15, 1056-18, 0812-09, 1038-07, 0814-10, 1041-19, 1170-04, 0929-08, 0919-16, 0814-03, 0926-11, 1035-01, 1034-19, 1035-10, 1174-19, 1176-03, 1067-02, 0919-17, 0920-05, 0927-13, 1060-17, 1060-05, 0927-17, 1044-18, 0812-10, 0811-20, 1041-20, 1174-04, 0925-05, 1060-01, 1062-09, 1066-20, 1174-07, 0814-15, 0814-17, 1053-16, 1054-02, 1034-02, 1052-16, 1052-19, 0918-06, 1052-13, 1052-20, 1052-15, 1034-07, 1043-10, 0811-15, 1054-03, 1061-08, 1043-06, 0926-10, 1044-06, 1034-03, 1052-09, 1043-12, 1055-01, 0929-09, 1038-05, 1034-01, 0810-19, 1054-20, 1062-07, 1059-20, 0814-16, 1043-14, 1062-08, 1054-01, 0929-07, 1038-06, 1056-19, 1057-03, 1169-20, 0926-19, 1067-01, 1038-04, 0925-06, 1043-08, 1043-13, 0812-11, 1057-05, 1060-06, 0926-16, and 1034-06.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJE
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register