Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Xomed, Inc.: Brand name: NIM EMG Endotracheal Tube REF 82-29106, 82-29107, 82-29108 NIM EMG Reinforced Endotracheal Tube REF 82-29306, 82-29307, 82-29308 [All products are packaged one unit per box.]

Recall number
Z-0310-05
Recalling firm
Medtronic Xomed, Inc.
Product
Brand name: NIM EMG Endotracheal Tube REF 82-29106, 82-29107, 82-29108 NIM EMG Reinforced Endotracheal Tube REF 82-29306, 82-29307, 82-29308 [All products are packaged one unit per box.]
Status
Terminated
Initiated
2004-10-29
Posted
2004-12-15
Terminated
2012-07-07
Reason
The firm received six reports of tubes with wires 'out of channel' two of which involved injuried patients. In one case in Germany it was reported that the patient had soreness of the throat and was discharged after four days. In the next case the patient had a tracheal wound. It was not clear what caused the wound.
Root cause
Other
Action
not on file
Quantity
22569 EMG Tubes
Distribution
Distributed nationwide and worldwide. to countries in the five continents with the highest populations.
Lot, serial or model codes
Product Code: 77 ETN  Lot Numbers: 32423300 through 35505400, and 35542200, 35542300, 35542400, 35542500, 35542600, 35543400, 35597200, 35602100, 35633400, 35633600, 35638200, 35638400
510(k) numbers
["K925640"]
PMA numbers
not on file
Product code
ETN
Firm FEI
1045254
Firm city
Jacksonville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register