Skip to the content

FDA Medical Device Recalls (openFDA)

Ameditech Inc: DrugSmart 11 Test Cup, Item No. 61127D; DrugSmart 12 Test Cup (Multi & BUP), Item No. 61210D; DrugSmart 9 Test Cup (BUP), Item No. 60960D; DrugSmart Cup 10, Item No. 61093D-3C These Drugs of Abuse Tests are one-step immunoassays for the qualitative detection of multiple drugs and drug metabolites in human urine. The configurations of these Drugs of Abuse Tests consist of a combination of drug strips, including Benzodiazepine (BZO). They are used to obtain a visual, qualitative result and are intended for professional and point of care use.

Recall number
Z-0308-2016
Recalling firm
Ameditech Inc
Product
DrugSmart 11 Test Cup, Item No. 61127D; DrugSmart 12 Test Cup (Multi & BUP), Item No. 61210D; DrugSmart 9 Test Cup (BUP), Item No. 60960D; DrugSmart Cup 10, Item No. 61093D-3C These Drugs of Abuse Tests are one-step immunoassays for the qualitative detection of multiple drugs and drug metabolites in human urine. The configurations of these Drugs of Abuse Tests consist of a combination of drug strips, including Benzodiazepine (BZO). They are used to obtain a visual, qualitative result and are intended for professional and point of care use.
Status
Terminated
Initiated
2015-10-20
Posted
not on file
Terminated
2016-11-09
Reason
Ameditech is recalling drug abuse tests because they have shown reduced reactivity.
Root cause
Nonconforming Material/Component
Action
Ameditech Inc. sent an Urgent Medical Device Recall Letter to al affected customers on October 20, 2015, to inform them that Ameditech is recalling affected lots of Ameditech's Drugs of Abuse Tests because they have shown reduced reactivity for up to three of the nineteen Benzodiazepine (BZO) compounds. The letter informs customers of the actions to be taken. Customers were instructed to complete and return the customer verification form to (858) 677-0243 or scan and email to amt.fieldaction@alere.com. For questions regarding this recall call 858-535-1968.
Quantity
1220 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia.
Lot, serial or model codes
140770 142294 151050 151527 142527 140998 142526 150596 151122 151196
510(k) numbers
["K052115","K113046","K113501"]
PMA numbers
not on file
Product code
JXM
Firm FEI
3003789989
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register