FDA Medical Device Recalls (openFDA)
Stelkast Co: Proven Cemented Semi-Constrained Total Knee. Part number SC1591. The size is 38 and the thickness is 10 mm. The Proven Knee System is a single use device intended for cemented reconstruction of the femoral and/or tibial portion of severely disabled and/or painful knee joints resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis with or without varus, valgus or flexion deformities, and revision surgery provided that there is radiographic evidence of sufficient sound bone to seat the prosthesis.
- Recall number
- Z-0308-2009
- Recalling firm
- Stelkast Co
- Product
- Proven Cemented Semi-Constrained Total Knee. Part number SC1591. The size is 38 and the thickness is 10 mm. The Proven Knee System is a single use device intended for cemented reconstruction of the femoral and/or tibial portion of severely disabled and/or painful knee joints resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis with or without varus, valgus or flexion deformities, and revision surgery provided that there is radiographic evidence of sufficient sound bone to seat the prosthesis.
- Status
- Terminated
- Initiated
- 2008-09-03
- Posted
- 2008-11-25
- Terminated
- 2008-12-04
- Reason
- Five (5) Three Peg Patella SC1591-29 were packaged in containers labeled for SC1591-38 sharing the same lot number.
- Root cause
- Employee error
- Action
- On 9/4/08, the recalling firm telephoned their customers to inform them of the problem and asked them to immediately recover all SC1591-38 with lot number 19906-052008. Once the product is in their possession, customers should contact Stelkast Customer Service for a Return Authorization (RA) Number prior to shipment to Stelkast.
- Quantity
- 5 units
- Distribution
- Nationwide Distribution --- including states of IA and WV.
- Lot, serial or model codes
- Part number SC1591-38. Lot number 19906-052008
- 510(k) numbers
- ["K980276"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 3004142400
- Firm city
- Mcmurray
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28