Atrium Medical Corporation: The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection chamber that holds shed mediastinal blood, a water seal chamber using a water seal that prevents air from entering into the pleural space and a suction control chamber, which incorporates a mechanical regulator to control suction pressures imposed upon the patient's pleural cavity.

Recall number
Z-0307-2022
Recalling firm
Atrium Medical Corporation
Product
The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection chamber that holds shed mediastinal blood, a water seal chamber using a water seal that prevents air from entering into the pleural space and a suction control chamber, which incorporates a mechanical regulator to control suction pressures imposed upon the patient's pleural cavity.
Status
Open, Classified
Initiated
2021-10-22
Posted
not on file
Terminated
not on file
Reason
Inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient
Root cause
Labeling design
Action
Atrium/Getinge issued Urgent Medical Device Correction letter issued October 22, 2021 via FedEx 2 day delivery. Letter states reason for recall, health risk and action to take: Our records indicate that you have received the Ocean Water Seal Chest Drain, Product Code/REF number 2002-400, having the lot numbers that are affected by this correction. Please ensure that all Ocean Water Seal Chest Drain, Product Code/REF number 2002-400, users at your facility are aware of this notice and post a copy of the Product Correction Notification on Page 4 in all inventory locations within your facility where the Atrium Ocean Chest Drains are stored. Please examine your inventory immediately to determine if you have any of the Ocean Water Seal Chest Drain, Product Code/REF number 2002-400, with the lot numbers listed in this notice. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL OCEAN CHEST DRAIN USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Atrium/Getinge by e-mailing a scanned copy to oceandrain.atrium@getinge.com or by faxing the form to (877) 448-0547. f you have any questions, please contact your Atrium/Getinge representative or call Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity
87,972 drains (14,662 cases; 6 units/case)
Distribution
US Nationwide Distribution Foreign: Bahrain Belgium Brunei Darussalam Canada Colombia Ecuador Finland Italy Kuwait Libya Mauritius Mexico Netherlands Poland Portugal Russian Federation Saudi Arabia Spain United Arab Emirates United Kingdom of Great Britain and Northern Ireland
Lot, serial or model codes
Product Code/REF Number 2002-400; Lot codes: 441533, 443300, 443608, 443629, 444473, 445527, 448966, 449683, 450899, 450900, 451270, 451988, 452177, 452814, 454577, 455871, 458264, 458604, 459198, 459642, 462192, 462194, 464075, 464076, 465037, 466247, 469921, 470627, 471651,471703, 472562, 472785 and 473725.  UDI Code: 00650862100211
510(k) numbers
["K043582"]
PMA numbers
not on file
Product code
KDQ
Firm FEI
3011175548
Firm city
Merrimack
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register