Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Elekta Leksell Gamma Plan Product Usage: Gamma Plan is a computer based system designed for Gamma Knife treatment planning.

Recall number
Z-0307-2012
Recalling firm
Elekta, Inc.
Product
Elekta Leksell Gamma Plan Product Usage: Gamma Plan is a computer based system designed for Gamma Knife treatment planning.
Status
Terminated
Initiated
2011-06-23
Posted
2011-11-30
Terminated
2012-01-31
Reason
There is a risk that the user may interpret the instructions in the manual too literally, and adjust the TMR 10 dose description to obtain an ETCPD that matches previous clinical prescriptions, which in turn could lead to a prescription dose that deviates a few percent from the intended dose.
Root cause
Labeling False and Misleading
Action
Elekta sent an IMPORTANT NOTICE letter dated June 23, 2011 to affected customers. The notice identified the product, problem, and corrective actions to be taken by the user. The letter advise all customers to follow any instructions or recommendations covered in the Notice. The letter states that Elekta is working on a new software version and customers will receive the future version when released. For questions contact your local Elekta representative.
Quantity
2
Distribution
Nationwide Distribution (USA) - including the states of: NM and WA
Lot, serial or model codes
6107 and 6115
510(k) numbers
["K973441"]
PMA numbers
not on file
Product code
IWB
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register