Philips Medical Systems (Cleveland) Inc: IQon Spectral CT (Model No. 728332), Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Recall number
Z-0306-2019
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
IQon Spectral CT (Model No. 728332), Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
Status
Terminated
Initiated
2017-06-07
Posted
not on file
Terminated
2019-07-03
Reason
It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.
Root cause
Software design
Action
The recalling firm sent Customer Information letters dated June 7, 2017, to affected customers. The letter notified customers of the product issue. Customers were instructed to determine whether they had a potentially affected system by checking the software version of their device. Customers were reminded that the "Impressions" section of the report is not intended for diagnostic use. Users should instead focus on the total calcium score shown in the "Findings" section, the coronary artery score(s) and/or the "Age and Gender" graph as identified in the Instructions for Use (IFU). The firm sent a follow-up dated July 3, 2018, to inform customers that software updates had been released which resolved the issue. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377 and follow the prompts).
Quantity
64
Distribution
US Nationwide Distribution
Lot, serial or model codes
software versions 4.7.0 or 4.7.2
510(k) numbers
["K171850"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register