Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Recall number
Z-0305-2012
Recalling firm
Elekta, Inc.
Product
Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Status
Terminated
Initiated
2011-03-09
Posted
2011-11-29
Terminated
2012-08-09
Reason
If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero value.
Root cause
Software design
Action
Elekta sent an IMPORTANT NOTICE letter dated March 9, 2011 to affected customers. The notice identified the product, the problem, and the corrective action to be taken by the customer. The notice advise all customers to follow any instructions or recommendations covered in the Notice. For questions call +44(0)1293 654200.
Quantity
56
Distribution
Nationwide Distribution (USA) including the states of: AZ, CA, GA, IL, LA, MD, MA, MI, MN, NJ, NY, NC, OH, PA, TN, VA, WI and Puerto Rico
Lot, serial or model codes
152019, 152448, 151076, 151954, 152454, 152117, 152452, 152393, 152394, 151942, 152163, 151955, 152176, 152222, 151931, 152388, 151141, 152365, 152303, 152358, 152353, 152359, 152352, 152362, 152305, 152468, 152481, 152469, 151812, 152432, 151260, 151952, 152370, 151396, 152301, 152248, 152425, 151155, 151156, 151157, 151412, 105984, 152307, 152308. 151443, 151545, 151628, 152207, 152271, 152373, 152371, 151080, 152211, 152158, 152392, 152331
510(k) numbers
["K051932"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register