FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-0305-2012
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Status
- Terminated
- Initiated
- 2011-03-09
- Posted
- 2011-11-29
- Terminated
- 2012-08-09
- Reason
- If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero value.
- Root cause
- Software design
- Action
- Elekta sent an IMPORTANT NOTICE letter dated March 9, 2011 to affected customers. The notice identified the product, the problem, and the corrective action to be taken by the customer. The notice advise all customers to follow any instructions or recommendations covered in the Notice. For questions call +44(0)1293 654200.
- Quantity
- 56
- Distribution
- Nationwide Distribution (USA) including the states of: AZ, CA, GA, IL, LA, MD, MA, MI, MN, NJ, NY, NC, OH, PA, TN, VA, WI and Puerto Rico
- Lot, serial or model codes
- 152019, 152448, 151076, 151954, 152454, 152117, 152452, 152393, 152394, 151942, 152163, 151955, 152176, 152222, 151931, 152388, 151141, 152365, 152303, 152358, 152353, 152359, 152352, 152362, 152305, 152468, 152481, 152469, 151812, 152432, 151260, 151952, 152370, 151396, 152301, 152248, 152425, 151155, 151156, 151157, 151412, 105984, 152307, 152308. 151443, 151545, 151628, 152207, 152271, 152373, 152371, 151080, 152211, 152158, 152392, 152331
- 510(k) numbers
- ["K051932"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28