Skip to the content

FDA Medical Device Recalls (openFDA)

Nova Ortho-Med Inc: 4205 Journey Rolling Walker Product Usage: used to assist people in walking

Recall number
Z-0304-2017
Recalling firm
Nova Ortho-Med Inc
Product
4205 Journey Rolling Walker Product Usage: used to assist people in walking
Status
Terminated
Initiated
2016-10-01
Posted
2016-10-24
Terminated
2017-02-06
Reason
Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.
Root cause
Component design/selection
Action
A customer notification letter dated 10/1/16 was sent to customers to inform them that the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out. This allows the frame support piece to detach from the walker and causes instability resulting in possible injury. The letter informs the customers of the root cause and the solution. Customers are instructed to record the serial #s of the product that have been corrected and email them to NOVA Customer Service (patricias@novajoy.com). Customers can also 1-800-557-6682 x131.
Quantity
1,380 units
Distribution
US Nationwide Distribution in the states of: CA, IN, HI, CT, UT, WA, FL, VA, WI, OH, MN, KS, MD, IL, MO, WY, MT, ME, PA, TN, MI, OR, NY, NJ, ID, NC, IA, OK, AR, VT, SC, TX
Lot, serial or model codes
Serial No. YA 4205RDFG004 TO YA4205RDFG0098 YZ4205BKFC0002 TO YZ4205RDFE0100
510(k) numbers
not on file
PMA numbers
not on file
Product code
ITJ
Firm FEI
3002329391
Firm city
Carson
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register