FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corporation: Battery Module for Sarns Modular Perfusion System 8000 Product Usage: The Sarns Modular Perfusion System 8000 is indicated for up to 6 hours use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist.
- Recall number
- Z-0303-2012
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Battery Module for Sarns Modular Perfusion System 8000 Product Usage: The Sarns Modular Perfusion System 8000 is indicated for up to 6 hours use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist.
- Status
- Terminated
- Initiated
- 2011-10-21
- Posted
- 2011-11-29
- Terminated
- 2012-02-17
- Reason
- Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled two connection between the board and the transformer.
- Root cause
- Employee error
- Action
- Terumo conducted a Correction/Removal - Customer Communicated Script on November 7, 2011 to the affected customer. The phone script communicated the correction and reason for removal, potential hazards, affected population, and confirmation of possession of affected module. A replacement date was set for November 7, 2011. For questions or concerns contact Terumo CVS Customer Service at 1-800-521-2818, hours are Monday to Friday. 8 AM -6 PM EST.
- Quantity
- 10 units
- Distribution
- Nationwide Distribution (USA) including the state of San Antonio, TX
- Lot, serial or model codes
- Catalog Number 16422 Serial Number 1901
- 510(k) numbers
- ["K915183"]
- PMA numbers
- not on file
- Product code
- DWB
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28