Skip to the content

FDA Medical Device Recalls (openFDA)

Ventana Medical Systems Inc: Ventana Medical System's Symphony Staining System, KPA, Model #: 2200000, Catalog #: 900-SYM1 & 900-SYM3, Ventana Medical Systems, Tucson, AZ

Recall number
Z-0300-2008
Recalling firm
Ventana Medical Systems Inc
Product
Ventana Medical System's Symphony Staining System, KPA, Model #: 2200000, Catalog #: 900-SYM1 & 900-SYM3, Ventana Medical Systems, Tucson, AZ
Status
Terminated
Initiated
2007-09-06
Posted
2008-01-09
Terminated
2008-07-08
Reason
Shock hazard: Devices released prior to full testing; the configuration of the Symphony instrument during Hi pot testing procedures did not completely evaluate all components in the test loop. In specific circumstances, a shock hazard exists.
Root cause
Process control
Action
On 9/6/07, the firm began visiting customers to perform the HiPot test. Prior to performing the High Potential Test of Symphony Instruments already in the Field, the Field Service Engineer or FSE ensures that the customer fully understands the purpose of the test and why it is necessary to perform the test at their site. The FSE conveys the following points regarding the HiPot test: - The HiPot test is a safety verification for the electrical path within the instrument. - The HiPot test was performed incorrectly during the manufacturing process. - The HiPot test is used to verify that if a high voltage overload condition were to occur it would not lead to a short/arc within the instrument, which could pose a risk to anyone that touches the instrument - The HiPot test places a voltage of 2,000VDC onto the high voltage path of the instrument and if any of the insulating components are compromised, the tester will notify the FSE of the condition. - If the instrument fails the HiPot test, the faulty component will be replaced and the instrument will be retested until it passes. - Due to the high voltage used the during the test, no one will be allowed to touch the instrument until the test is complete. The FSE performing the test documents that the customer was informed of the reason for the High Potential Test on their Field Service Report.
Quantity
45 units
Distribution
Worldwide: USA, Japan and France
Lot, serial or model codes
Serial Numbers: 910006-910011, 910013-910016, 910019, 910020, 910022, 910023, 910029, 910031-910033, 910036, 910049, 910053, 910054, 910056, 910057, and 910060.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KEY
Firm FEI
1000125183
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register