FDA Medical Device Recalls (openFDA)
Smiths Medical PM, Inc.: BCI 3404 Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor. Smiths Medical PM, Inc. This device is designed to provide full featured monitoring capabilities in a tabletop design. The system features an ECG cable interface, an SpO2 probe interface, display of patient and waveform data via an EL panel, power status LED, and the function keypad area consisting of six keys (on/off, waveform/trend, alarm silence, menu/enter, up arrow & down arrow). The monitor has a serial port that is used for data communications to a printer or computer and for analog outputs.
- Recall number
- Z-0300-2007
- Recalling firm
- Smiths Medical PM, Inc.
- Product
- BCI 3404 Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor. Smiths Medical PM, Inc. This device is designed to provide full featured monitoring capabilities in a tabletop design. The system features an ECG cable interface, an SpO2 probe interface, display of patient and waveform data via an EL panel, power status LED, and the function keypad area consisting of six keys (on/off, waveform/trend, alarm silence, menu/enter, up arrow & down arrow). The monitor has a serial port that is used for data communications to a printer or computer and for analog outputs.
- Status
- Terminated
- Initiated
- 2006-11-06
- Posted
- 2006-12-19
- Terminated
- 2008-10-09
- Reason
- An error occurred at the board manufacturing site which may affect the following parameters: - An electrical noise observed on the SpO2 pleth waveform generating and displaying erratic heart rates. % SpO2 reading displays dashes. - The ECG waveform may appear noisy, ECG readings appear accurate. -Respiration waveform may appear noisy and the respiration readings will be displayed as dashes.
- Root cause
- Other
- Action
- An Safety Action Bulletin, dated 11/03/2006 was sent to affected customers describing the issue, what parameters are affected on the monitor and provides informatin on how to have the monitor repaired. Smiths Medical MDPM requests that an acknowledgement form be returned to them.
- Quantity
- 207
- Distribution
- Worldwide, including USA, Puerto Rico, Argentina, Bolivia, Brunei, Chile, China, Columbia, Czech Republic, Germany, Greece, Hong Kong, India, Israel, Kuwait, Mexico, Netherlands, Poland, Romania, Singapore, Spain, Taiwan, and United Kingdom.
- Lot, serial or model codes
- Catalog Number 3404001, serial number ranges: AI05070075 - AI05070099, AI05090000 - AI05090024, AI05110040 - AI05110059, AI05120000 - AI05120010, AI05120013 - AI05120024, AI06010035 - AI06010049, AI06040000 - AI06040024, AI06060000 - AI06060014, AI06060016 - AI06060024, AI06070035 - AI06070041, AI06070043 - AI06070045. Catalog Number 3401001S, serial number range: AI05110000 - AI05110039
- 510(k) numbers
- ["K981939"]
- PMA numbers
- not on file
- Product code
- DQA
- Firm FEI
- 2183502
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28