Terumo Cardiovascular Systems Corporation: The Terumo Advanced Perfusion System 1, Catalog Number 801764 System 1 Base, 220 V Product Usage: The Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.

Recall number
Z-0299-2012
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
The Terumo Advanced Perfusion System 1, Catalog Number 801764 System 1 Base, 220 V Product Usage: The Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.
Status
Terminated
Initiated
2011-11-08
Posted
2011-11-29
Terminated
2012-08-22
Reason
1 US customer was operating the Advanced Perfusion System 1 with an unapproved, modified cable, which was in product development, to connect the system an an occluder for the Sarns Modular Perfusion System 8000 . Five(5)of twelve(12) cables from Terumo Japan, which were in product development, were given to a US customer.
Root cause
Component design/selection
Action
Terumo sent an URGENT MEDICAL DEVICE CORRECTION: Recall Notification dated November 8, 2011 to all affected customers. The letter described the reason for the recall, known/potential hazards and request immediate location and correction of product. The letter instructs customers to stop using the modified cables and return to Terumo for replacement and destruction. Customers were asked to complete the attached response form and fax to the number indicated on the form. All affected customers will be receiving a safety advisory to advise them that Terumo do not recommend the alternative configuration of a System 1 Base with an 8K Occluder. For questions or concerns contact Terumo CVS Customer Service at 1-800-521-2818 or 1-800-292-6551 (fax).
Quantity
12 units
Distribution
Worldwide Distribution - (USA) Nationwide Distribution
Lot, serial or model codes
Lot numbers: 0006-1415; Catalog # 801764 System 1 Base, 220 V
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register