Skip to the content

FDA Medical Device Recalls (openFDA)

Advanced Bionics Corporation: Advanced Bionics Precision Linear Leads, part of the Precision Spinal Cord Stimulation System. Model Numbers: SC-2138-30, SC-2138-50, SC-2138-50T and SC-2138-70.

Recall number
Z-0299-2007
Recalling firm
Advanced Bionics Corporation
Product
Advanced Bionics Precision Linear Leads, part of the Precision Spinal Cord Stimulation System. Model Numbers: SC-2138-30, SC-2138-50, SC-2138-50T and SC-2138-70.
Status
Terminated
Initiated
2006-09-21
Posted
2007-01-10
Terminated
2012-01-25
Reason
A small number of unimplanted Precision Linear Leads may have been assembled with incorrect electrode material. At high stimulation levels, the metal may corrode and dissolved metals may enter the patient.
Root cause
Other
Action
In United States, the clinician recall notification letters were distributed by Fed-Ex or U.S. Postal Service certified mail on 09/21/2006. For the other regions, the in-country representatives started providing notification to the clinicians on 09/21/2006. The recall notification letters will instruct physicians to immediately return the unimplanted units.
Quantity
18
Distribution
Worldwide.
Lot, serial or model codes
Serial Numbers: 111744, 112122, 112141, 112224, 112233, 114114, 116547, 116553, 117860, 117907, 110490, 100600, 110625, 110809, 113745, 113780, 112055 and 118943
510(k) numbers
not on file
PMA numbers
["P030017"]
Product code
LGW
Firm FEI
1000220815
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register