Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Renal Div: Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.

Recall number
Z-0299-05
Recalling firm
Baxter Healthcare Renal Div
Product
Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.
Status
Terminated
Initiated
2004-12-02
Posted
2004-12-08
Terminated
2006-07-21
Reason
The instrument power cord may become disconnected from the instrument and pose a fire or electrical shock hazard.
Root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 12/2/04 to all Meridian customers on the same date. The letters informed the accounts of the potential for disconnection of the power cord and teh potential for arcing, fire or electrical shock. The accounts were requested to examine their instruments to see if a new power cord clamp has been installed on the unit, using the enclosed instructions and complete the enclosed reply sheet indicating the number of units already corrected, and the number of units still requiring the power cord clamp. The accounts were requested to fax the completed reply for to Baxter at 847-270-5457.
Quantity
3,004 units
Distribution
Nationwide and internationally to Mexico, China and Korea.
Lot, serial or model codes
serial numbers 200101 to 202190
510(k) numbers
["K992894"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register