FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: Vitros Immunodiagnostic Products CEA REAGENT PACK, Lot 750, REF 192 0115, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
- Recall number
- Z-0299-04
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- Vitros Immunodiagnostic Products CEA REAGENT PACK, Lot 750, REF 192 0115, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
- Status
- Terminated
- Initiated
- 2003-12-03
- Posted
- 2004-01-06
- Terminated
- 2004-02-20
- Reason
- Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
- Root cause
- Other
- Action
- U.S. customers notified via FED EX letter sent on 12/3/03. Letter includes modified QC procedures to use until replacement product is issued. Foreign affiliates notified via e-mail on 12/3/03 with instructions to notify customers.
- Quantity
- 524 reagent packs
- Distribution
- 154 reagent packs shipped to consignees in the U.S. 370 reagent packs shipped to Ortho/J&J International Affiliates for further distribution to foreign users. Foreign consignees are located in Argentina, Australia, Belgium, Brazil, Chile, China, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Portugal, Puerto Rico, Singapore, Spain, United Kingdom, and Venezuela.
- Lot, serial or model codes
- Lot #750, Exp. 14 January 2004
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- DHX
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28