St Jude Medical CRMD: St. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)
- Recall number
- Z-0298-2007
- Recalling firm
- St Jude Medical CRMD
- Product
- St. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)
- Status
- Terminated
- Initiated
- 2006-10-06
- Posted
- 2006-12-19
- Terminated
- 2012-03-26
- Reason
- St. Jude Medical has identified a low-frequency anomaly in the software used in the APS III Model 3500/3510 and Merlin PCS Model 3650 programmers that can lead to incorrect reporting of battery voltage, expected battery longevity and Elective Replacement Indicator (ERI) status.
- Root cause
- Other
- Action
- The notification material was sent to physicians by FedEx. The advisory notification was sent by FedEx on Oct. 12, 2006, to arrive at the physicians' offices on Oct. 13th. In the US, acknowledgement of receipt of the notification material will be monitored by St. Jude Medical CRMD in Sylmar, California. For the rest of the world it will be by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on October 13, 2006. Plus information concerning this advisory can be found on St. Jude's website, www.sim.com Completion of this advisory is estimated to be three months after its initiation.
- Quantity
- 199,065
- Distribution
- Worldwide in USA, ARGENTINA, AUSTRALIA, AUSTRIA, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, ESTONIA, EUROPE, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GUATEMALA, HONG KONG, ICELAND, INDIA, ISRAEL, ITALY, JAPAN, JORDAN, SOUTH KOREA, LATVIA, LICHTENSTEIN, LUXEMBURG, MALAYSIA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, PORTUGAL, SAUDI ARABIA, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, UNITED KINGDOM & VENEZUELA
- Lot, serial or model codes
- All Devices (all serial numbers for the devices referenced in the advisory.)
- 510(k) numbers
- not on file
- PMA numbers
- ["P880086"]
- Product code
- DXY
- Firm FEI
- 2017865
- Firm city
- Sylmar
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28